Significant Errors In Insulin Dose Can Result When Blood Glucose Meters Are Miscoded According to New Clinical Study

Accurate Coding is Key to Preventing Potentially Serious Health Complications Associated with Insulin Overdose


TARRYTOWN, N.Y., Nov. 6 (HSMN NewsFeed) -- When persons with diabetes use miscoded blood glucose meters to determine how much insulin to take, significant errors in insulin dose can result that may potentially lead to short- and long-term health complications, according to findings of a new study presented at the Sixth Annual Diabetes Technology Meeting in Atlanta, Georgia.

The American Diabetes Association estimates that there are 14.6 million children and adults diagnosed with diabetes in the United States(1), of which an estimated 4.4 million, or 30%, require insulin to manage their disease(2). Those who require insulin must closely monitor their blood sugar with a blood glucose meter to plan their meals, exercise regimens and insulin dosage.

In this study, for certain miscoded meters, the probability of insulin error of plus or minus 2 units of insulin was 50% as compared to 8% for correctly, manually coded meters. The probability of insulin dose error of plus or minus 3 units of insulin was 23% for the miscoded meters but only 0.5% for the manually correctly coded meters.

Coding is the process by which a blood glucose meter is matched to each new box of test strips being used. This is done either by inserting a code strip or code chip into the meter, or by entering a code number into the meter. If this step is not performed, the meter may give inaccurate results leading to wrong therapy. For example, relying on a miscoded blood glucose meter to determine how much insulin to take can result in a potentially harmful overdose. Insulin overdose may cause dangerously low blood sugar (hypoglycemia) leading to behavioral changes, confusion, loss of consciousness and, if untreated, seizure, coma and even death. Chronic under-dosing of insulin may contribute to the long-term health problems associated with high blood sugar including kidney disease, nerve disease, eye problems, and heart disease.

"When dealing with patients with diabetes we've observed that many either do not understand what proper coding is, or do not realize its importance. Patients sometimes use expired test strips and/or fail to properly code their blood glucose meters to the lot of test strips they are using," said Dr. Steven Edelman, an author on the study and Professor of Medicine, division of Endocrinology and Metabolism, University of California, San Diego, and the Veterans Affairs Medical Center and founder of TAKING CONTROL OF YOUR DIABETES (www.tcoyd.org).

The study findings also showed that auto-coded meters (meters that automatically set the correct code anytime a test strip is inserted) gave more accurate blood glucose values than meters that had been correctly coded manually. This also translated into a lower probability of insulin dose error. For auto-coded meters, the probability of plus or minus 1 unit and plus or minus 2 units of insulin could be as high as 35.4% and 1.4% respectively. However, with the auto-coded meters, there were no calculated insulin dose errors above plus or minus 2 units.

"These findings are significant because studies have shown that approximately 16% - or one out of six - persons failed to properly manually code their blood glucose meters to the lot of test strips being used(3). Understanding the potentially serious consequences of relying on a meter that is not properly coded - is essential for every person with diabetes, especially those who need to take insulin," said Linda Schrock, a nurse and certified diabetes educator, who was also an investigator on the study, at Elkhart General Hospital, Elkhart, Indiana.

The study authors concluded that to avoid insulin dosing errors, people should be carefully instructed how to correctly code their meters or be advised to use an auto-coded meter.

Study Design

The study involved 116 patients at three clinical centers. The blood glucose values for patients in this study ranged from 52 - 498 mg/dL. After fasting, the patients were given a two-hour meal tolerance test. At zero, 60 and 120 minutes the study subjects' fingerstick blood was tested on five different popular blood glucose meters (two were auto-coded meters). Some of the meters were purposely miscoded to the lot of test strips. The auto-coded meters were always properly coded due to their inherent design. The values from all the meters were compared with blood glucose values measured on a laboratory glucose analyzer to determine how accurate (inaccurate) the meters were.

Glucose values obtained from some of the miscoded meters used for this study showed an average error ranging between plus 29% and minus 37%.

A Monte Carlo simulation, (a statistical method that uses existing data sets to forecast performance in the field) was conducted on the data from the clinical trial to generate 'ideal' and 'simulated-meter' glucose values, and subsequent insulin doses. This simulation was based on various assumptions such as, one unit (1U) of insulin covers 50mg/dL blood glucose.* From these calculations, the probability of insulin dose errors for the three types of blood glucose meters (miscoded, manually correctly coded and autocoded) were determined.

The probability for an error of plus or minus one unit of insulin was 44.6% for correctly coded meters compared to 49.6% for incorrectly coded meters. The probability for an error with a miscoded meter of plus or minus four units of insulin was 2.8% and for plus or minus five units of insulin was 0.06%. There was no instance of a plus or minus four or five unit error with correctly, manually coded meters. For auto-coded meters there were no calculated insulin dose errors above plus or minus two units.

Bayer HealthCare, Diabetes Care

Bayer HealthCare, Diabetes Care is a worldwide leader in diabetes, supporting customers in 100 countries. Since the introduction of CLINITEST® reagent tablets in 1941, Bayer has led the way in diabetes care product innovation. The company changed the face of diabetes care in 1969 when it introduced the first portable blood glucose meter and test strips. Bayer HealthCare further innovated diabetes management by being the first company to introduce a suite of blood glucose monitors that do not require coding. The BREEZE® and CONTOUR® blood glucose monitoring systems offer people with diabetes an unparalleled choice in diabetes management systems. Recently, the Arthritis Foundation in the United States and the Arthritis Society of Canada each granted Ease-of Use Commendation to the BREEZE meter, representing the first time a blood glucose meter has been recognized as easy to use for arthritis sufferers.

In July 2006, Bayer Diabetes Care acquired Metrika Inc., maker and manufacturer of A1CNow+®, a meter-based diabetes monitoring system for measurement of HbA1c (glycated hemoglobin) an important indicator of long term blood sugar control.

Bayer HealthCare, Diabetes Care global headquarters is located in Tarrytown, New York, in the United States and operates as part of Bayer HealthCare LLC, a member of the worldwide Bayer HealthCare group.

Bayer HealthCare AG

Bayer HealthCare, a subsidiary of Bayer AG, is one of the world's leading, innovative companies in the health care and medical products industry based in Leverkusen/Germany. In 2005, the Bayer HealthCare subgroup generated sales amounting to some 9.4 billion Euro. Bayer HealthCare employed 33,800 people worldwide in 2005.

The company combines the global activities of the divisions Animal Health, Consumer Care, Diabetes Care, Diagnostics and Pharmaceuticals. Since January 1, 2006 the new Pharmaceutical Division consists of the former Biological Products and Pharmaceutical Division and now comprises three business units: Hematology/Cardiology, Oncology and Primary Care.

Bayer HealthCare's aim is to discover and manufacture products that will improve human and animal health worldwide. The products enhance well-being and quality of life by diagnosing, preventing and treating diseases.

* Small variations may occur due to the nature of Monte Carlo simulation.
(1) American Diabetes Association:
www.diabetes.org/diabetes-statistics.jsp
(2) Roper 2005 U.S. Diabetes Patient Marker Study, April 19, 2006
(3) Raine, C.H. Endo Prac 9: pg 137, 2003

This news release contains forward-looking statements based on current assumptions and forecasts made by Bayer Group management. Various known and unknown risks, uncertainties and other factors could lead to material differences between the actual future results, financial situation, development or performance of the company and the estimates given here. These factors include those discussed in our public reports filed with the Frankfurt Stock Exchange and with the U.S. Securities and Exchange Commission (including our Form 20-F). The company assumes no liability whatsoever to update these forward-looking statements or to conform them to future events or developments.


Source: Bayer HealthCare